The DriftconRF hardware
|
What is DriftconRF ? |
|
The DRIFTCON® RF system is a construction in which the thermistors are positioned in a holder, which has PCR plate dimensions.
This "plate" also contains all processing electronics, calibration values, power cells, as well as RF communication electronics. While cycling, all temperature values are transmitted to the supplied RF USB dongle connected to the computer. Usage of the system is monitored by means of software and the DRIFTCON® RF intelligent hardware.
DRIFTCON® and DRIFTCON® RF test results are analyzed and reported through an automated process, thereby minimizing user input interventions and prohibiting any alteration of the acquired sample data.
|
|
End-user product |
|
Driftcon® RF is an end-user product.
Use is recommended daily, weekly or monthly.
Measurements are performed statically as well as dynamically
Measurements on all (q) PCR closed systems non accessible by regular Driftcon systems, regardless brand or type
Measurements are related to the ITS-90 Scale
Measurements and test systems are NIST traceable
Accuracy is better than ±0.20°C over a range of 10 - 100°C Measurement range 5-100°C
Measurements of 15 channels simultaneously (2 Hz/channel)
Retrospective analysis, and movie player in place
Tests reports of drift, uniformity and accuracy, heat rate, cool rate, plateau time
Access to global user database
Instrument to instrument protocol transition tool
Conform GMP, GLP, ISO, Ster-Lab, governmental and association regulations
Recalibration every 120 measurements, usage counter on board, or annually
|
|
DriftconRF system components |
|
DRIFTCON® RF intelligent fixture, electronics and transmitter incorporated in the plate, includes also lid measurement sensors
DRIFTCON® USB dongle, and extension cable
DRIFTCON® RF fixture charger
DRIFTCON® RF software
Instrument software specification set (market specs)
Membership of the CYCLERtest global threshold database
DRIFTCON® system storage suitcase
Optional:
Instrument software specification sets for other types/brands thermal instruments
Notebook with DRIFTCON® RF software pre-installed
|
|
IMPORTANT |
|
Compatibility :
Applied Biosystems 7300 Real-Time PCR System
Applied Biosystems 7500 Real-Time PCR System
Applied Biosystems 7900 Fast Real-Time PCR System with standard 96-Well Block Module
Stratagene MX3000P, MX3005P
Eppendorf Mastercycler ep realplex
The DriftconRF® system addresses a new market segment that can not be served with a standard DRIFTCON® system. The use of the DriftconRF® system is only recommended in above mentioned instruments. Nearly all other types of instruments can be measured using the standard DRIFTCON® systems. A measurement with a DriftconRF® system in other types of instruments can lead to unexpected results!
Calibration: Annually or after 120 validations. (whatever comes first)
Configurations :
96V-15: 15 temperature sensors
16-6: 16 temperature sensors, all well format (available soon)
96V-15LP: 15 temperature sensors, low profile format (available soon)
16-6LP: 16 temperature sensors , all well, low profile format (available soon)
384-15: 15 temperature sensors for 384 block format (available soon)
Warranty:
This high precison instrument is specifically designed for usage in above mentioned Real-Time Cyclers and is sensitive to damage if used in other instruments. Warranty is voided when DRIFTCON® RF is used in other instruments.
|
|
|
|